9841AS Nitrogen
Oxides Analyzer
The Teledyne ML 9841AS Nitrogen Oxides Analyzes uses gas-phase chemiluminescence detection to perform continuous analysis of nitric oxide (NOx), total
oxides of nitrogen (NOx) and nitrogen dioxides (NOx).
The Teledyne ML 9841AS analyzer design represents an advance in nitrogen
oxides analysis technology archived primarily by using adaptive microprocessor control of
a single measurement channel. The instrument consists of a pneumatic system, an NO2 to NO
converter (MOLYCON), a reaction cell, photomultiplier tube and processing electronics.
The analysis of nitrogen oxides by means of chemiluminesence is
generally acknowledged to be in the best direct measurement technique. The method is based
on the luminescence of an activated molecular NO2 species produced by the reaction between
NO and O3 in an evacuated chamber. The NO molecules react with ozone to form the activated
species, NO2, according to the reaction mechanism:
NO + O3 à NO2 * + O2
As NO2* reverts to \a lower energy state, it emits broad-band
radiation form 500 to 3,000 nm, with a maximum intensity at approximately 1,100nm Since on
NO molecule is required to form one NO2* molecule, the intensity of the chemiluminesence
reaction is directly proportional to the NO concentration in the sample.
Note that the chemiluminesence reaction only occurs between O3
and NO. Consequently, analysis of NO2 to NO conversion is accomplished with a catalytic
converter, the MOLYCON, that pneumatically switches in and out of the sample stream in six
second cycles. This switching creates a time-phased sample stream composed of NO and NOx
samples. The NOx cycle measures both the NO and the NO2 in the sample by use of the
MOLYCOM, while the NO cycle measures only the NO. The NO2 concentration is derived by
subtracting the NO signal from the Nox2 signal.
FEATURES:
High range NO measurement capabilities. The
Teledyne ML 9841AS is
specially designed for systems applications with a range of 0-1000 ppm. Please see
specifications for more detailed information.
Because the
Teledyne ML 9841AS has no chopper, there is no need to
replaces belts or motors.
* Internal data logging capability. Optional internal floppy disk
data storage provides additional data backup.
|
|
|