1.    Customer Contact Information

 

*Name:
Title:
*Company:
*Phone:
FAX:
Email:
*Stack Identification:

 

 

 

 

 

 

2.   *Mounting Flange  (SELECT ONE OF THE FOLLOWING)

 

I will supply my own 2 inch flanges for the Standard Installation

I wish to use something other than 2 inch flanges, described below:

 

3.   *Flange to Flange Distance  (COMPLETE ONE OF THE FOLLOWING)  

Feet     Inches

or

Meters

4.    Temperatures  (SELECT ONE OF THE FOLLOWING FOR EACH GROUP)

A.  *PROCESS

Less than 500 Degrees Fahrenheit (290 Degrees Celsius)
501 to 1500 Degrees Fahrenheit (261 to 815 Degrees Celsius)
less than 1500 Degrees Fahrenheit (less than 815 Degrees Celsius) Consult Factory

 

B.  *AMBIENT (STACK)

-25 to 140 Degrees Fahrenheit (-32 to +60 Degrees Celsius)
Other, Consult Factory
 

C.  AMBIENT (REMOTE)

+32 to 140 Degrees Fahrenheit (0 to 40 Degrees Celsius)
Other, Consult Factory

 

5.    *PROCESS PRESSURE  (SELECT ONE OF THE FOLLOWING)

 

-15.0 to 5.0inches H2O (-3.745 to +1.245 kilopascal)
-15.0 to 15.0inches H2O (-3.745 to +3.745 kilopascal)
Other, Consult Factory

 

 

6.    MOUNTING CLEARANCE  (SELECT ONE OF THE FOLLOWING)

The standard Mounting Clearance is sufficient.
I will use something other than the standard Mounting Clearance, described below:
 

 

7.    PURGE FAIL CABLE LENGTH  (SELECT ONE OF THE FOLLOWING)

 

The standard Purge Fail Cable Length of 40 ft. is sufficient.
I will use something other than 40 ft. of Purge Fail Cable, described below:
 

 

8.    DATA CABLE LENGTH  (The Data Cable is supplied at an additional cost.)

 

I wish to purchase: Feet of Data Cable.

 

 

 

 

9.    STACK POWER CIRCUIT REQUIREMENTS

 

Optical Head: 85-245 VAC, 47-63Hertz, Single Phase, 30 VA Maximum
Single Purge Blower Number: 115/230 VAC, Single Phase, 414 VA Maximum
Dual Purge Blower Number: Two circuits, each with the same requirements as a Single Purge Blower
Number (#): Does not apply to instrument Air Purge Version

 

 

10.    *CORRECTION FOR EXIT PATHLENGTH


 

Monitor Pathlength: Feet Inches
Outlet Pathlength (round): Feet Inches

or

Rectangular Exit Width:  Feet Inches
Rectangular Length: Feet Inches

or

Desired Pathlength Correction Factor:
   
If you are requesting a specific Pathlength Correction Factor value, please explain:

 

11.    *MAXIMUM OPACITY LIMIT

 

Instantaneous Opacity Limit   Average Limit
less than 20% Opacity less than 20% Opacity
greater than or =20% Opacity greater than or =20% Opacity

 

 

12.    SPECIFY FULL SCALE

 

Opacity:

Optical Density:
Particulate Mass:

 

 

13.    USE AUTO CAL SEQUENCE?  (SELECT ONE OF THE FOLLOWING)

YES
NO

                    

 

14.     SYSTEM CONFIGURATION  (SELECT ONE OF THE FOLLOWING)

 

Direct Interface
Enhanced Remote Panel

        

 

If you selected the Direct Interface Configuration, fill out the following:

Output Units: Opacity Optical Density Particulate Mass, Consult Factory
Current Outputs: 4 to 20 milliamp 0 to 20 milliamp
Relay Output #1:
Relay Output #2:
Digital Input #1:
Digital Input #2:

 

 

Only if you selected the Enhanced Remote Panel Configuration, fill out this page:

 

Analog Output #1:
Analog Output #2:
Analog Output #3:
Analog Output #4:

 

 

Relay K1:
Relay K2:
Relay K3:
Relay K4:
Relay K5:
Relay K6:
Relay K7:
Relay K8:

 


Ethernet Module Setup

 

Use a Dynamic Internet Provider (IP) address
Use a Static Internet Provider (IP) address, if yes please supply the information below:

Static IP Address  
Subnet Mask  
Default Gateway

 

Enhanced Remote Panel Power Circuit Requirements:

 

85-245 Volts Alternating Current, 47-63 Hertz, Single Phase, 30 Volts Alternating Current Maximum
115 Volts Alternating Current Power Cord 230 Volts Alternating Current Power Cord

Confirmation:   Validation Enter Confirmation Code:   *
 


How to complete the LightHawk® 560 Installation Checklist

1.    CUSTOMER CONTACT INFORMATION:

Fill out your name, phone number, email address and other information. The Stack Identification should be the name of the stack or duct that the monitor will be on. Some sites may have a name such as “Boiler1” others may have names such as “West Stack” or a more descriptive name of the process.

2.    MOUNTING FLANGES (Not typically provided)

The Standard mounting flanges are 2" 150lb ANSI flanges and are to be supplied by the customer. The flanges are welded to 2" Sch. 40 pipe that is also customer supplied. If the application calls for mounting pipes longer than 13", larger diameter pipes should be considered in order to maintain good alignment (consult drawing 1800-0013).  If you are adapting to an EXISTING flange or using mounting pipes other than 2" Sch. 40 , select Other and consult the TML factory to specify the configuration.

3.    FLANGE TO FLANGE CALIBRATION DISTANCE

Record this distance in English or Metric units. This distance is to be measured from the outside of one mounting flange to the outside of the other mounting flange with an accuracy of +\-1% of the distance. (consult drawing 1800-0007) Teledyne Monitor Labs recommends that the user measure this dimension after the mounting pipe and flanges have been installed and consider using a laser tape device to make the measurement.
*NOTE: The accuracy of the factory calibration and opacity measurement are both dependent on this dimension being measured and reported accurately.

4.    TEMPERATURES

  • Select the Process, Ambient Stack and Ambient Remote temperature Ranges.

  • Select the Typical process temperature under normal operating conditions.

This information is necessary to ensure that the hardware configuration chosen is adequate for the user's application (high temperature seals, stainless steel nozzles and a heat shield may be required when the temperature exceeds 500 ºF)

5.    PROCESS PRESSURE
Select the process static pressure range in INCHES of WATER at the monitor installation location.

This information is necessary to ensure that the hardware configuration chosen is adequate for the user's application (high positive pressure >5" H2O may require options)

6.    MOUNTING CLEARANCES

Review the proposed installation locations and reference the Teledyne Monitor Labs drawings. Check OK if sufficient clearances are available. Mark Other and consult the factory if alternative installation configurations are required. 

  • Stack Equipment: Be certain that there is sufficient clearance between the stack mounted equipment and any potential obstructions (i.e.: distance to floor, horizontal distance to catwalk rail). Weather hood is…20" W x 22 ¼" L x 33" H. (See drawing 1800-0007).

7.    PURGE FAIL CABLE LENGTH

The purge fail cable is supplied with the system and connects the Retroreflector with the Optical Head assembly. This cable must be routed through customer-supplied 1/2" (min.) flexible and rigid conduit. A Standard 40 feet (12.2meters) of purge fail cable will be provided with each monitor purchased. List cable lengths that exceed 40 feet under Other.

Consult the TML factory for pricing of cables greater than 40 feet in length.

8.    DATA CABLE LENGTH

DATA CABLE LENGTH Direct Interface (Not provided with standard system, available as option)

The cost of the cables from the Optical Head to the output or display device has not been included in the Continuous Omissions Monitoring Systems base price. TML will provide the cable at additional cost to the customer.

The user must determine the length of the data cable measured from the Optical Head to their DAS/PLC or other output device. Minimum recommended conduit size is 1/2". A Direct Interface configuration requires two cables; a non-shielded 4-conductor/20 AWG (Alpha 66-44) and a 6-shielded pair/24 awg (Alpha 6386/Belden 9991). Consult the Teledyne Monitor Labs factory for current price and delivery of data cable.

DATA CABLE LENGTH with Enhanced Remote (Not provided with standard system, available as option) The cost of the cable from the Optical Head to the output or display device has not been included in the Continuous Omissions Monitoring Systems base price. TML can provide the cable at additional cost to the customer.

The user must determine the length of the data cable measured from the Optical Head to the Enhanced Remote Panel. Minimum recommended conduit size is 1/2". The Enhanced Remote Panel configuration requires only a single twisted pair, 16 AWG with shield (Alpha 5610B1601).  Consult the Teledyne Monitor Labs factory for current price and delivery of data cable.

9.    STACK POWER CIRCUIT REQUIREMENTS

Listed are the supply voltage and current capacity of the customer supplied circuits for the STACK equipment. Circuits must meet the following minimum requirements.

Stack Equipment: Maximum power draw is:

  • 14VA single phase for the Optical Head

  • 414VA single phase for one Purge Blower

  • 828VA single phase for dual Purge Blowers

A minimum of one 115VAC single phase, 10A circuit is required. The customer is to hardwire 115VAC into the purge motor junction box of both the Retroreflector and Optical Head sides if a dual purge system is utilized.

10.    CORRECTION FOR STACK EXIT PATHLENGTH

Most regulatory agencies require that monitor opacity readings be corrected to reflect the observable opacity where the emissions exit to the atmosphere. The Environmental Protection Agency (EPA) calls this the Path Length Correction Factor (PLCF).

Record the following to calculate the PathLength Correction Factor (PLCF):

  • a.) Monitor Pathlength

Record the inside dimension of duct or stack at the monitor location.

  • b.) Emission Pathlength

  • For a round stack - record the inside diameter of duct or stack at the exit

  • For square or rectangular duct or stack, record both dimensions at the exit.

11.    MAXIMUM OPACITY LIMIT

Changes to US EPA regulations require manufacturers to configure and test Continuous Opacity Monitoring Systems (COMs) differently based on the user's maximum opacity emissions. Record the opacity limits imposed by your controlling regulatory agency.

If the user's operating permit includes both Instantaneous and Integrated opacity limits, please select both limits and the Integration period on the checklist.

12.    SPECIFY FULL SCALE

The Standard output scaling includes:

Opacity = 100%
Optical Density = 2.000
Particulate Mass = Consult the Teledyne Monitor Labs factory

The user's operating permit may require a specific opacity full scale based on the classification of the emission source.

13.    USE AUTO CAL SEQUENCE

Check YES if the Auto Cal sequence is to be used to collect calibration data. Check NO if the user will be exclusively using External Mode control.

  • Auto Cal Sequence - The real time clock in the Optical Head determines the start and duration of each phase of the calibration data.

14.    USE EXTERNAL MODE CONTROL

Select Direct Interface or Enhanced Remote configuration according to application.

15.    SPECIFY CAL CYCLE OUTPUT TIMES:

Calibration information is usually provided on Channel 2. The standard auto cal reporting cycle is 120-seconds for each of the following phases: ZERO, UPSCALE (240-Sec total). PLCF and DUST values are not output in the standard configuration. Check YES to accept the Standard 120-seconds for ZERO and Upscale. If NO is selected, please specify on the checklist each phase duration you want. Each is independently programmable. Maximum allowable time is 2,880 seconds (max. 720 secs. each).

16.    SYSTEM CONFIGURATION:

(Select one) Direct Interface or Enhanced Remote
The LightHawk® 560 is available in two basic configurations. The Direct Interface configuration does not have the Enhanced Remote Panel. The Direct Interface requires more wires to be run from the stack or duct. The Enhanced Remote option gives the user more information down at the ‘remote’ location and offers more inputs and outputs than the Direct Interface. The Enhanced Remote option requires only one cable with two wires to be run from the stack or duct to the Enhanced Remote Panel’s location.

This is option A, Direct Interface

17A.    SPECIFY OUTPUT UNITS:

Check the analog output units your monitor is to report in, Opacity, Optical Density or Particulate Mass (select one) Standard default = Opacity

18A.    CURRENT OUTPUTS:

The Direct Interface has two analog outputs available. They are isolated from each other and circuit common.

Channel 1 is always dedicated to the Instantaneous value and does not carry calibration values.  Channel 2 is always Selectable Average with calibration. If an output carries calibration values (Channel 2), the analog output will track the calibration values when it is commanded to go into calibration or when it is scheduled to go into calibration. If the output does not carry calibration values (Channel 1), the analog output will hold the last NORMAL mode value when the monitor is in a calibration mode.

Please select the output current configuration you want, 4 - 20 ma or 0 - 20 ma. Select only one.  Standard default = 4 - 20ma.

19A.    RELAY OUTPUTS:

The Six Point I/O board uses 2 relays that can be configured to actuate on various output modes or monitor conditions. Select any two from the table below. These functions are set via jumper settings and can be easily modified at the site. Contacts are SPDT Form C. The two relays available can be configured as either Normally Open or Normally Closed via a jumper setting.  The standard selections most commonly chosen are “Calibration Data on Analog Outputs” and “Instrument Malfunction”. The following table lists the available selections.

Function

Calibration Data on Analog Outputs

Instrument Malfunction

Instantaneous Alarm 1 Actuated

Selectable Average Alarm 1 Actuated

UPSCALE data on Analog Outputs

ZERO data on Analog Outputs

Purge Failure

Excessive Dirt Compensation

20A.    DIGITAL INPUT SELECTION

Digital Inputs
Two discrete inputs are provided to allow monitor output mode control by user supplied circuits.  These optical isolators are activated by either dry contact closures or +5
Voltage Direct Current signal from a user-supplied circuit. These signals are transmitted to the Optical Head vi a the Data cable.  This configuration is field selectable and may be modified at the time of instrument start up The following definitions apply to the selection of the External Mode control digital inputs:

Force Upscale = The instrument goes into calibration UPSCALE mode. The calibration mechanism moves to the UPSCALE position. Analog output #1 holds the last stack value read, analog output #2 follows the UPSCALE value

Force Zero = The instrument goes into calibration ZERO mode. The calibration mechanism moves to the ZERO position. Analog output #1 holds the last stack value read analog output #2 follows the ZERO value.

Dump PLCF = The calibration mechanism moves to NORMAL position. Real time stack data appears on the LED readout and on analog output #1. Analog output #2 follows the Pathlength Correction Factor  value.

Dump Dust = The calibration mechanism moves to NORMAL position. Real time stack data appears on the LED readout and on analog output #1. Analog output #2 follows the DUST COMPENSATION value.

Force Cal Cycle = The calibration mechanism moves to ZERO, UPSCALE, Pathlength Correction Factor , DUST COMPENSATION and then back to NORMAL. It remains in each mode for a configurable period of time.

This is Option B, Enhanced Remote Option

17B. SPECIFY DESIRED ANALOG OUTPUTS:

The Standard Configuration includes:

Channel 1          Instantaneous Opacity, with cal values displayed.
Channel 2          Instantaneous Opacity, without cal values displayed.
Channel 3          Pathlength Correction Factor
Channel 4          Dust Compensation

The Analog Outputs can be easily changed at a later date, through a list on the Enhanced Remote Panel. Choose the desired Analog Outputs from the following table:

Parameters for Analog Output 

Parameter Name

Functional description

Instant Opacity

Instantaneous opacity corrected for the system’s Pathlength Correction Factor . Update time is 6-8 seconds.

Minute Ave Opacity

A one minute average of the instantaneous opacity reading. This value is Pathlength Correction Factor  corrected. The period is minute to minute according to the real time clock in Optical Head.

Selectable Ave Opacity

An average opacity reading whose time period is selectable in one-minute increments. The time selected should divide evenly into 60 minutes. 1,2,4,6,10,12,15,30 and 60-minute averages are possible. Six minutes is the standard selection.

Instant Opt Density

Instantaneous optical density corrected for the system’s Pathlength Correction Factor. Update time seconds. This is a mathematical conversion of the opacity reading.

Minute Ave Opt Density

A one minute average of the instantaneous optical density reading. This value is Pathlength Correction Factor  corrected. The period is minute to minute according to the real time clock in the Optical Head.

Select Ave Opt Density

An average Optical Density reading whose time period is selectable in one-minute increments. The time selected should divide evenly into 60 minutes. 1, 2, 4, 12, 15, 30, and 60-minute averages are possible. Six minutes is the standard selection.

Instant Particulate Mass

Instantaneous Particulate Mass loading in mg/m 3 . Update time is 6-8 seconds.  This is a mathematical correlation from the opacity reading by a curve that is developed through extractive testing. This is not available for output until the correlation testing is done and the results entered into the 560’s software.

Minute Ave Particulate Mass

A one minute average of the instantaneous Particulate Mass loading reading.  period is minute to minute according to the real time clock in the Optical Head.  available for output until testing results are entered.

Select Ave Particulate Mass

 

An average Particulate Mass loading reading whose time period is selectable one-minute increments. The time selected should divide evenly into 60 minutes.  2, 4, 6, 10, 12, 15, 30, and 60-minute averages are possible. Six minutes is standard selection. Not available for output until testing results are entered.

Dust Compensation Output

The amount of Dust Compensation build up . The Dust Compensation factor applied to the NORMAL and UPSCALE mode readings is limited to ±4% PLC corrected opacity. However, the Dust Compensation reading is unbounded.

Zero Output

The calibration Zero value.

Upscale Output

The calibration Upscale value.

PLCF Output

The Path Length Correction Factor being applied to the readings.


18B. RELAY CONTACT OUTPUTS

There are eight relays (K1-K8) available on the LightHawk® 560 Multi I/O Module to interface with user circuits. Each relay is SPST with one normally open contact rated 3A max. The Standard Configuration includes:

K1        Analog outputs displaying calibration ZERO
K2        Analog outputs displaying UPSCALE calibration
K3        Pathlength Correction Factor value being displayed on the analog outputs
K4        Analog outputs displaying any type of calibration data
K5        Data Valid (Failsafe)
K6        Instantaneous Opacity exceeds Level 1 set-point
K7        Average Opacity exceeds Level 1 set-point
K8        Purge Fail

The relays can also be easily changed at a later date, through a list on the Enhanced Remote Panel. Select options for the relay configurations from the following table:

Parameters for Relay Assignment
(Digital Output closure conditions)

Opacity
Exceedence

Optical Density
Exceedence

Mass Loading
Exceedence

System
Trouble

Calibration
Control

Instantaneous Opacity
Level #1 

Instantaneous Opacity Density
Level #1 

Instantaneous Mass load
Level #1 

Cal Failure

Normal on Average Opacity

Instantaneous Opacity
Level #2

Instantaneous Opacity Density
Level #2

Instantaneous Mass load
Level #2

Excess Dust
Compensation Comp

Cal on Average Opacity

Min Opacity
Level #1

Min Opt Density
Level #1

Min Mass load
Level #1

Fault

Dust
Compensation Average Opacity

Min Opacity Level #2

Min Opt Density
Level #2

Min Mass load
Level #2

Data Valid

PLCF on

Average Opacity

Ave Opacity Level #1

Ave Opt Density
Level #1

Ave Mass load
Level #1

Purge Failure

Zero on

Average Opacity

Ave Opacity Level #2

Ave Opt Density
Level #2

Ave Mass load
Level #2

No Selection

Upscale on

Average Opacity

19B.    DIGITAL INPUTS

The Enhanced Remote has 8 isolated inputs available. The location and designation of each are fixed. To use any of these inputs, merely connect wires to the appropriate input. A simple jumper setting on the Multi- I/O board determines if they are activated by a dry contact or +5 Voltage Direct Current input.  Dry contact closure is the standard configuration. More information is provided on the Installation Drawings and in the Manual.

Input #1            Force Upscale
Input #2            Force Zero
Input #3            Dump Pathlength Correction Factor
Input #4            Dump Dust Comp
Input #5            Force Cal Cycle
Input #6            Not Used
Input #7            Test Zero Scale
Input #8            Test Full Scale

20B. ENHANCED REMOTE PANEL POWER CIRCUIT CAPACITY

Maximum power draw is:
25 VA single phase for the remote equipment.

The Enhanced Remote Panel normally operates on a 115VAC, 60 Hertz, single phase supply. The assembly is usually shipped with a standard straight 115VAC, 3-pronged molded power cord, 15feet