Model T300M with NumaView™ Software 

Mid-Range Gas Filter Correlation Carbon Monoxide Analyzer​

T300M Product Sheet

TML T-300M NEW.png

Description

The Model T300M CO analyzer measures low to mid ranges of carbon monoxide by comparing infrared energy absorbed by a sample to that absorbed by a reference gas according to the Beer-Lambert law.  Using a Gas Filter Correlation Wheel, a high-energy IR light source is alternately passed through a CO filled chamber and a chamber with no CO present.  The light path then travels through the sample cell, which has a folded path of 2.56 meters.

The energy loss through the sample cell is compared with the span reference signal provided by the filter wheel to produce a signal proportional to concentration, with little effect from interfering gases within the sample.  This design produces excellent zero and span stability and high signal to noise ratio, allowing excellent performance from low to high range.

All T Series instruments offer an advanced color display, capacitive touch screen, intuitive user interface, flexible I/O, and built-in data acquisition capability.  All instrument set up, control and access to stored data and diagnostic information is available through the front panel, or via RS232, Ethernet, or USB com ports, either locally or by remote connection.

The Model T300M comes with NumaView™ Software. NumaView™ Remote PC Software allows for a remote connection with virtual interface and data downloading capability to analyzers operating NumaView™ Software. ​​


Downloads

​Tit​le​​​​​
​File ​​Size
T300M Ma​nual
7767KB
​T300M Specifications​
​1252KB
​T Seri​es Br​ochure
​1030​K​B​

Features

  • Ranges: 0-5 ppm to 0-5,000 ppm, user selectable
  • Dual ranges
  • Large, vivid, and durable color graphics display with touch screen interface
  • Ethernet, RS-232, and (optional) USB com ports
  • Front panel USB connections for peripheral devices and firmware upgrades
  • 8 analog inputs (optional)
  • Adaptive signal filtering optimizes response time
  • Temperature & pressure compensation
  • Comprehensive internal data logging with programmable averaging periods
  • Ability to log virtually any operating parameter
  • Two-year warranty
  • Five-year warranty on GFC wheel​


Specifications

Ranges Min: 0 - 5 ppm full scale
Max: 0 - 5,000 ppm full scale (selectable, dual-range supported)
Measurement Units  ppm, μg/m3, mg/m3 (selectable)
Zero Noise < 0.1 ppb (RMS)
Span Noise < 0.5% of reading (RMS) above 20 ppm
Lower Detectable Limit  0.2 ppm
Zero Drift < 0.5 ppm/24 hours
Span Drift < 0.5% of full scale/24 hours
Lag Time  10 seconds
Rise/Fall Time < 60 seconds to 95%
Linearity ±1% of full scale
Precision  1% of reading
Sample Flow Rate  800 cc/min ±10%
Power Requirements 100V-120V, 220V-240V, 50/60 Hz
Analog Output Ranges 10V, 5V, 1V, 0.1V (selectable)
Recorder Offset ±10%
Included I/O 1 x Ethernet: 10/100Base-T
2 x RS232 (300-115,200 baud)
2 x USB device ports
8 x opto-isolated digital outputs
6 x opto-isolated digital inputs
4 x analog outputs
Optional I/O 1 x USB com port
1 x RS485
8 x analog inputs (0-10V, 12-bit)
4 x digital alarm outputs
Multidrop RS232
3 x 4 - 20mA current outputs
Operating Temperature Range 5 - 40ºC
Dimensions (HxWxD) 7” x 17” x 23.5” (178 x 432 x 597 mm)
Weight

Specifications are subject to change without notice.
All specifications are based on constant conditions.​

 40 lbs (18.1 kg)
          


Maintenance

The following table provides a list of items that  need regular maintenance to ensure your product remains in good working condition.

ITEMACTIONFREQUENCY
Particulate filter element ReplaceWeekly or as needed
Pump diaphragmsReplaceAnnually
Sample flowVerifyCheck for proper flow (800cc/min ± 20%) annually
Dark CalCalibrate/AdjustPerform annually as needed, or after adjusting R7 on Sync-Demod  board.
Pneumatic linesExamine and cleanAs necessary
Leak CheckVerify leaksLeak Check after maintenance or repair and at least annually
Clean optical benchCleanAs needed

  ​​