Model T200 with NumaView™ Software

​Chemiluminescence Oxides of Nitrogen Analyzer

T200 Product Sheet

TML T-200 NEW.png 

Description

The Model T200 NO/NO2/NOx analyzer uses the proven chemiluminescence detection principle, coupled with state-of-the-art electronics to allow accurate and dependable low level measurements for use as an ambient analyzer or dilution CEMS monitor.

A unique AutoZero feature provides superb stability by continuously correcting for zero drift, while advanced Adaptive Filtering allows the analyzer to optimize performance under changing conditions.  The Model T200 includes a permeation inlet dryer for ozone generation to provide excellent reliability with no periodic replacement required.  A catalytic exhaust ozone scrubber is standard for maximum safety and pump life.

All T Series instruments offer an advanced color display, capacitive touch screen, intuitive user interface, flexible I/O, and built-in data acquisition capability.  All instrument set up, control and access to stored data and diagnostic information is available through the front panel, or via RS232, Ethernet, or USB com ports,​ either locally or by remote connection.

The Model T200 comes with NumaView™ Software. NumaView™ Remote PC Software allows for a remote connection with virtual interface and data downloading capability to analyzers operating NumaView™ Software​.​


Downloads

​Title​​​
​F​ile Size
​T200 Manual
​10911KB
​T200 Specifications
822KB
T Seri​es ​Brochure​
​1030​K​​B​

Features

  • Ranges: 0-50 ppb to 0-20 ppm, user selectable
  • Selectable ranges, with dual range capability
  • Large, vivid, and durable color graphics display with touch screen interface
  • Ethernet, RS-232, and (optional) USB com ports
  • Front panel USB connections for peripheral devices and firmware upgrades
  • 8 analog inputs (optional)
  • Adaptive signal filtering optimizes response time
  • Temperature & pressure compensation
  • Internal zero & span check (optional)
  • Comprehensive internal data logging with programmable averaging periods
  • Ability to log virtually any operating paramete
  • Two-year warranty​


Specifications

Ranges Min: 0 - 50 ppb full scale
Max: 0 - 20,000 ppb full scale (selectable, dual-range supported)
Measurement Units ppb, ppm, μg/m3, mg/m3 (selectable)
Zero Noise < 0.2 ppb (RMS)
Span Noise < 0.5% of reading (RMS) above 50 ppb
Lower Detectable Limit 0.4 ppb
Zero Drift < 0.5 ppb/24 hours
Span Drift < 0.5% of full scale/24 hours
Lag Time 20 seconds
Rise/Fall Time < 60 seconds to 95%
Linearity 1% of full scale
Precision 0.5% of reading above 50 ppb
Sample Flow Rate 500 cc/min ±10%
Power Requirements 100V-120V, 220V-240V, 50/60 Hz
Analog Output Ranges 10V, 5V, 1V, 0.1V (selectable)
Recorder Offset ±10%
Included I/O 1 x Ethernet: 10/100Base-T
2 x RS232 (300-115,200 baud)
2 x USB device ports
8 x opto-isolated digital outputs
6 x opto-isolated digital inputs
4 x analog outputs
Optional I/O 1 x USB com port
1 x RS485
8 x analog inputs (0-10V, 12-bit)
4 x digital alarm outputs
Multidrop RS232
3 x 4 - 20mA current outputs
Operating Temperature Range 5 - 40ºC (with US EPA Approval)
Dimensions (HxWxD) 7” x 17” x 23.5” (178 x 432 x 597 mm)
Weight Analyzer: 40 lbs (18 kg)
External pump: 15 lbs (7 kg)
Certifications​





Specifications are subject to change without notice.
All specifications are based on constant conditions.​

US EPA: RFNA-1194-099
EU: EN14211 TÜV Rheinland
QAL1 Certified: EN15267
MCerts: Sira MC050068/11
CNEMC: 质(认)字 No. 2015-028 Report ​​​




Maintenance

​​

​The following table provides a list of items that  need regular maintenance to ensure your product remains in good working condition. ​ 

ITEMACTIONFrequencyCal Check Required
TEST functionsReview and evaluateWeeklyNo
Particulate filterChange particle filterWeekly (if in stack system: As Needed)No
Zero/span checkEvaluate offset and slopeWeeklyNo
Zero/span calibrationZero and span calibrationEvery 3 monthsYes
External zero air scrubber optionExchange chemicalEvery 3 monthsNo
External dryer optionReplace chemicalWhen indicator color changesNo
Ozone cleanserChange chemicalAnnuallyYes
Reaction cell window (“optical filter”  Figure 11‑6 in manual)CleanAnnually or as necessaryYes
DFU filtersChange particle filterAnnuallyNo
Pneumatic sub-systemCheck for leaks in gas flow pathsAnnually or after repairs involving pneumaticsYes if a leak is repaired
Reaction cell O-rings & sintered filtersReplaceAnnuallyYes 
PMT Sensor Hardware CalibrationLow-level hardware calibrationOn PMT/ preamp changes or if slope is outside of 1.0±0.3Yes
PumpRebuild headwhen RCEL pressure exceeds 10 in-Hg-A (at sea level)Yes
Inline Exhaust ScrubberReplaceAnnuallyNo
NO2 converterReplace converterEvery 3 years or if conversion efficiency drops below 96%Yes 
Desiccant  bagsReplaceAny time PMT housing is opened for maintenancen/a