Model T200H with NumaView™ Software

High Range Chemiluminescence Oxides of Nitrogen Analyzer

T200H Product Sheet

TML T-200H NEW.png 

Description

The Model T200H combines the proven chemiluminescence principle with a number of innovative features to provide reliable and repeatable NO, NO2, and NOx measurements for extractive type CEM systems.  A high-efficiency stainless steel thermal NO2 converter provides durable and consistent operation under the harshest conditions (other converter options also available).

The T200H may be fitted with an optional, internal paramagnetic O2 sensor or an infrared absorption CO2 sensor, reducing integration and operating costs significantly. 

All T Series instruments offer an advanced color display, capacitive touch screen, intuitive user interface, flexible I/O, and built-in data acquisition capability.  All instrument set up, control and access to stored data and diagnostic information is available through the front panel, or via RS232, Ethernet, or USB com ports, either locally or by remote connection.

The Model T200H comes  with NumaView™ Software. NumaView™ Remote PC Software allows for a remote connection with virtual interface and data downloading capability to analyzers operating NumaView™ Software. ​


Downloads

​Tit​le​​​​​
​File ​​Size
​T200H Ma​nual
​7762KB
​T200H Specifications​
​933KB
​T Seri​es Br​ochure
​1030​K​B​

Features

  • Ranges: 0-5 ppm to 0-5,000 ppm, user selectable
  • Independent ranges
  • Optional internal O2 or CO2 sensor
  • Large, vivid, and durable color graphics display with touch screen interface
  • Ethernet, RS-232, and (optional) USB com ports
  • Front panel USB connections for peripheral devices and firmware upgrades
  • 8 analog inputs (optional)
  • Adaptive signal filtering optimizes response time
  • Temperature & pressure compensation
  • Comprehensive internal data logging with programmable averaging periods
  • Ability to log virtually any operating parameter
  • Two-year warranty


Specifications

Ranges​​ Min: 0 - 5 ppm full scale
Max: 0 - 5,000 ppm full scale
(selectable, independent NO, NO2, NOX ranges with dual range supported)
Measurement Units ppm, mg/m3 (selectable)
Zero Noise < 20 ppb (RMS)
Span Noise < 0.2% of reading (RMS) above 20 ppm
Lower Detectable Limit 40 ppb
Zero Drift < 20 ppb/24 hours
Span Drift < 0.5% of reading/24 hours
Lag Time 20 seconds
Rise/Fall Time < 60 seconds to 95% (in switching mode)
Linearity 1% of full scale
Precision 0.5% of reading above 5 ppm
Sample Flow Rate 290 cc/min ±10%
Power Requirements 100V-120V, 220V-240V, 50/60 Hz
Analog Output Ranges 10V, 5V, 1V, 0.1V (selectable)
Recorder Offset ±10%
Included I/O 1 x Ethernet: 10/100Base-T
2 x RS232 (300-115,200 baud)
2 x USB device ports
8 x opto-isolated digital outputs
6 x opto-isolated digital inputs
4 x analog outputs
Optional I/O 1 x USB com port
1 x RS485
8 x analog inputs (0-10V, 12-bit)
4 x digital alarm outputs
Multidrop RS232
3 x 4-20mA current outputs
Operating Temperature Range 5 - 40ºC
Dimensions (HxWxD) 7” x 17” x 23.5” (178 x 432 x 597 mm)
Weight



Specifications are subject to change without notice.
All specifications are based on constant conditions.​

Analyzer: 40 lbs (18 kg)
External pump: 15 lbs (7 kg) ​



Maintenance

​The following table provides a list of items that  need regular maintenance to ensure your product remains in good working condition.

ITEMACTIONFREQUENCYCAL CHECKMANUAL SECTION
Particulate FilterChange filterWeekly (or as needed)No6.3.1
Verify Test FunctionsReview and evaluateWeeklyNo6.2; Appendix C in manual
Zero/Span CheckEvaluate offset and slopeWeekly-- 5.4, 5.6, 5.8
Zero/Span CalibrationZero and span calibrationEvery 3 months-- 5.2, 5.5, 5.8, 5.9
NO2 ConverterReplace converter & check efficiencyEvery 3 years or if conversion efficiency <90%Yes if CE factor is used--
External, 2-stage Zero Air Scrubber (Optional)Exchange chemicalEvery 3 months  (or as needed)No3.5.3.2
Reaction Cell WindowClean optics, Change O-ringsAnnually or as necessaryYes6.3.5
Air Inlet Filter Of Permeation DryerChange particle filterAnnuallyNo6.3.1, 6.3.2 
Pneumatic Sub-SystemCheck for leaks in gas flow pathsAnnually or after repairs involving pneumaticsYes on leaks, else no7.5.1, 7.5.2
All Critical Flow Orifice O-Rings & Sintered FiltersReplaceAnnuallyYes 6.3.6
1 PumpRebuild headAnnuallyYes6.3.3.1
Inline Exhaust ScrubberReplaceAnnuallyNo--
PMT Sensor Hardware CalibrationLow-level hardware calibrationOn PMT/ preamp changes & if 0.7< slope >1.3Yes7.6.4

 

1 A pump rebuild kit is available from Teledyne API's Technical Support, including all instructions and required parts.

 

  ​