Model T802 with NumaView™ Software

​Paramagnetic Oxygen Analyzer

T802 Product Sheet

TML T-802 NEW.png 

Description

The Model T802 O2 uses the proven paramagnetic measurement principle, coupled with state of the art microprocessor technology, to provide accurate and dependable measurements of Percent-Level O2.

Paramagnetic technology is non-depleting, which means that there are no consumable parts, and ensures a consistent performance over time.  The selectivity of the paramagnetic measurement for oxygen means that there is little or no interference from other common background gases.  The T802 offers a stable and inherently linear measurement of oxygen, making it possible to calibrate the analyzer by checking only two points.  There is no requirement for a reference gas during operation.

All T Series instruments offer an advanced color display, capacitive touch screen, intuitive user interface, flexible I/O, and built-in data acquisition capability.  All instrument set up, control and access to stored data and diagnostic information is available through the front panel, or via RS232, Ethernet, or USB com ports, either locally or by remote connection.

The Model T802 comes with the NumaView™ Software. NumaView™ Remote PC Software allows for a remote connection with virtual interface and data downloading capability to analyzers operating NumaView™ Software. ​


Downloads

​Tit​le​​​​​
​File ​​Size
​T802 Ma​nual
5854KB
​T802 Specifications​
​1097KB
​T Seri​es Br​ochure
​1030​K​B​

Features

  • O2 Range: 0-1% to 0-100%, user selectable
  • Dual ranges and auto ranging
  • Large, vivid, and durable color graphics display with touch screen interface
  • Ethernet, RS-232, and (optional) USB com ports
  • Front panel USB connections for peripheral devices and firmware upgrades
  • 8 analog inputs (optional)
  • Adaptive signal filtering optimizes response time
  • Temperature & pressure compensation
  • Comprehensive internal data logging with programmable averaging periods
  • Ability to log virtually any operating parameter
  • Two-year warranty

​ 


Specifications


Ranges Min: 0 - 1 % full scale
Max: 0 - 0-100% F​ull scale (selectable
Zero Noise < 0.2 % ppb (RMS)
Span Noise < 0.05% of reading (RMS)
Lower Detectable Limit 0.04 %
Zero Drift < ±0.02% /24 hours; < ±0.05% /7 days
Span Drift < ±0.1% /7 days
​Temperature Coefficient < ±0.01% per degree C
Accuracy < ±0.1%
Rise/Fall Time < 60 seconds to 95%
Linearity < ±0.1%
Precision 0.5% of reading above 50 ppb
Flow Rate  120 ml ± 20ml/min
​Humidity Range
​0 - 95% RH
​​     Pressure Range
​25-31 in Hg
Power Requirements 100V-120V, 220V-240V, 50/60 Hz
Analog Output Ranges 10V, 5V, 1V, 0.1V (selectable)
Recorder Offset ±10%
Included I/O 1 x Ethernet: 10/100Base-T
2 x RS232 (300-115,200 baud)
2 x USB device ports
8 x opto-isolated digital outputs
6 x opto-isolated digital inputs
4 x analog outputs
Optional I/O 1 x USB com port
1 x RS485
8 x analog inputs (0-10V, 12-bit)
4 x digital alarm outputs
Multidrop RS232
3 x 4 - 20mA current outputs
Operating Temperature Range 5 - 40ºC
Dimensions (HxWxD) 7” x 17” x 23.5” (178 x 432 x 597 mm)
Weight  28 lbs (12.7 kg) ​​
 
 
 
 
 
 
 
 



Applications

Maintenance

​The following table provides a list of items that  need regular maintenance to ensure your product remains in good working condition.

Item

Action

Frequency

Cal Check Required

Manual Section

Particulate Filter

Replace

Weekly or as needed

No

10.3.1

Verify Test Functions

Record and analyze

Weekly or after any Maintenance or Repair

No

11.1.2

Pump Diaphragm

Replace

Annually

Yes

10.3.2

Perform Flow Check

Check Flow

Annually

No

10.3.4

Perform Leak Check

Verify Leak Tight

Annually or after any Maintenance or Repair

No

10.3.3

Pneumatic lines

Examine and clean

As needed

Yes if cleaned

 

Chassis

Wipe down

As needed

Only if cover removed