Sulfur Compounds
​​
​​​​​​​​​​​​​​​​​​​​​​​​ Conta​​​ct us
Traini​ng
Sales Contacts
Service Contacts
Terms & Conditions

Model T100 with NumaView™ Software​

UV Fluorescence Sulfur Dioxide Analyzer​

T100 Product Sheet

TML T-100 NEW.png

 Downloads

​Title​​​
​File Size
T100 Manual
​8386KB
T100 Specifications
​1277KB
​T Series Brochure
​1030​KB

 Description

​​

The Model T100 uses the proven UV fluorescence principle, coupled with a state of the art user interface to provide easy, accurate, and dependable measurements of low level SO2.

Exceptional stability is achieved with the use of an optical shutter to compensate for PMT drift and a reference detector to correct for changes in UV lamp intensity.  A hydrocarbon ‘kicker’ and advanced optical design combine to prevent inaccuracies due to interferents.

All T Series instruments offer an advanced color display, capacitive touch screen, intuitive user interface, flexible I/O, and built-in data acquisition capability.  All instrument set up, control and access to stored data and diagnostic information is available through the front panel using NumaView™ Software, or via RS232, Ethernet, or USB com ports, either locally or by remote connection using the included NumaView™ Remote PC Software. NumaView™ Remote PC Software allows for a remote connection with virtual interface and data downloading capability to analyzers operating NumaView™ Software.

The Model T100 comes with NumaView ™ Software. NumaView™ Remote PC Software allows for a remote connection with virtual interface and data downloading capability to analyzers operating NumaView ™ Software. ​

 Features

​​
  • Ranges: 0-50 ppb to 0-20 ppm, user selectable
  • Selectable ranges, with dual range capability
  • Large, vivid, and durable color graphics display with touch screen interface
  • Ethernet, RS-232, and (optional) USB com ports
  • Front panel USB connections for peripheral devices and firmware upgrades
  • 8 analog inputs (optional)
  • Adaptive signal filtering optimizes response time
  • Temperature & pressure compensation
  • Internal zero & span check (optional)
  • Comprehensive internal data logging with programmable averaging periods
  • Ability to log virtually any operating parameter
  • ​Two-year warranty

 Specifications

Ranges
Min: 0 - 50 ppb full scale
Max: 0 - 20,000 ppb full scale (selectable, dual-range supported)
Measurement Units ppb, ppm, μg/m3, mg/m3 (selectable)
Zero Noise < 0.2 ppb (RMS)
Span Noise < 0.5% of reading (RMS) above 50 ppb
Lower Detectable Limit 0.4 ppb
Zero Drift < 0.5 ppb/24 hours
Span Drift < 0.5% of full scale/24 hours
Lag Time 20 seconds
Rise/Fall Time < 100 seconds to 95%
Linearity 1% of full scale
Precision 0.5% of reading above 50 ppb
Sample Flow Rate 650 cc/min ±10%
Power Requirements 100V-120V, 220V-240V, 50/60 Hz
Analog Output Ranges 10V, 5V, 1V, 0.1V (selectable)
Recorder Offset ±10%
Included I/O 1 x Ethernet: 10/100Base-T
2 x RS232 (300-115,200 baud)
2 x USB device ports
8 x opto-isolated digital outputs
6 x opto-isolated digital inputs
4 x analog outputs
Optional I/O 1 x USB com port
1 x RS485
8 x analog inputs (0-10V, 12-bit)
4 x digital alarm outputs
Multidrop RS232
3 x 4 - 20mA current outputs
Operating Temperature Range 5 - 40ºC (with US EPA Approval)
Dimensions (HxWxD) 7” x 17” x 23.5” (178 x 432 x 597 mm)
Weight 35.7 lbs (16.2 kg)
Certifications*






Specifications are subject to change without notice.
All specifications are based on constant conditions.​
US EPA: EQSA-0495-100
EU: EN14212 TÜV Rheinland
QAL1 Certified: EN15267
MCerts: Sira MC0500067/07
CNEMC: 质(认)字 No. 2015-028 Report ​​




 Maintenance

​The following table provides a list of items that  need regular maintenance to ensure your product remains in good working condition.

ITEMACTIONFREQUENCYCAL CHECKMANUAL SECTION
1Particulate filterChange particle filterWeeklyNo10.3.1
Verify test functionsReview and evaluateWeeklyNo10.2; Appendix C  in manual
Zero/span checkEvaluate offset and slopeWeekly-- 9.3, 9.6, 9.9
1Zero/span calibrationZero and span calibrationEvery 3 months-- 9.2, 9.4, 9.5, 9.7, 9.8
1External zero air scrubber (optional)Exchange chemicalEvery 3 monthsNo10.3.3
1Perform flow checkCheck FlowEvery 6 MonthsNo10.3.7
Internal IZS Permeation TubeReplaceAnnuallyYes10.3.2
Perform pneumatic leak checkVerify Leak TightAnnually or after repairs involving pneumaticsYes10.3.6
2Pump diaphragmReplaceAnnuallyYesRefer to diaphragm kit instructions
Calibrate UV Lamp OutputPerform lAMP​ CAL Prior to zero/span calibration or PMT hardware calibration-- 5.9.6 & 11.7.2.5
3PMT sensor hardware calibrationLow-level hardware calibrationOn PMT/ preamp changes  if 0.7 < slope or slope >1.3Yes11.7.2.8
1Sample chamber optics Clean chamber, windows and filtersAs necessaryYes11.7.2.2 & 11.7.2.3
1Critical flow orifice & sintered filtersReplaceAs necessaryYes10.3.4

 

 

1 These items are required to maintain full warranty; all other items are strongly recommended.

2 A pump rebuild kit is available from Teledyne API's Technical Support, including all instructions and required parts

 3 Replace desiccant bags each time the inspection plate for the sensor assembly is removed.​