The Model T300M CO analyzer measures low to mid ranges of carbon
monoxide by comparing infrared energy absorbed by a sample to that
absorbed by a reference gas according to the Beer-Lambert law. Using a
Gas Filter Correlation Wheel, a high-energy IR light source is
alternately passed through a CO filled chamber and a chamber with no CO
present. The light path then travels through the sample cell, which has
a folded path of 2.56 meters.
The energy loss through the sample
cell is compared with the span reference signal provided by the filter
wheel to produce a signal proportional to concentration, with little
effect from interfering gases within the sample. This design produces
excellent zero and span stability and high signal to noise ratio,
allowing excellent performance from low to high range.
All T
Series instruments offer an advanced color display, capacitive touch
screen, intuitive user interface, flexible I/O, and built-in data
acquisition capability. All instrument set up, control and access to
stored data and diagnostic information is available through the front
panel, or via RS232, Ethernet, or USB com ports, either locally or by
remote connection.
The Model T300M comes with NumaView™ Software.
NumaView™ Remote PC Software allows for a remote connection with virtual
interface and data downloading capability to analyzers operating
NumaView™ Software.