Model N​300
Gas Filter Correlation CO Analyzer
with NumaView™ Softwar​​​​​e


Description

​The Model N300 is a microprocessor-controlled analyzer based on the Beer-Lambert law that is used to determine the concentration of carbon monoxide (CO) in a sample gas.

The N300 uses Gas Filter Correlation (GFC) to overcome the interfering effects of various other gases (such as water vapor) that also absorb Infrared (IR) light. The analyzer passes the IR light beam through a spinning wheel made up of two separate chambers: one containing a high concentration of CO known as the reference, and the other containing a neutral gas known as the measure. The concentration of CO in the sample chamber is computed by taking the ratio of the instantaneous measure and reference values and then compensating the ratio for sample temperature and pressure.

Instrument functions and controls are managed through a series of integrated microprocessor-controlled modules utilizing a simple and reliable CAN Bus communications architecture. Each module is independently assembled and calibrated allowing easy and fast field replacement to maximize instrument uptime. The long-life sample filter option further improves efficiency with a ~6 months exchange interval in ambient air quality monitoring applications.

Intuitive operation and calibration of all N Series products is achieved through the NumaView™ Software interface. The graphical user interface (GUI) is customizable, giving the user fast and efficient access to instrument status, as well as measurement data and diagnostic parameters in either numeric or graphical form. NumaView™ Remote Software (included at no charge) provides the same virtual interface and complete instrument control, as well as access to the instrument's large internal data storage buff​er from a remote PC or tablet.​

Features

  • Single or dual range capability
  • Customizable alerts and continuous self-checking
  • Wide operating temperature range
  • Internal DC-powered vacuum pump
  • 5-year warranty on Gas Filter Correlation (GFC) wheel
  • Integrated Oxygen (O2) or Carbon Dioxide (CO2) sensor (optional)
  • Internal zero and span valves (optional)
  • Optional 47mm membrane​ or long-life​ sample particulate filter


Specifications

 RangesMin: 0 - 1 ppm full scale
Max: 0 - 1,000 ppm full scale (selectable, dual-range supported)
 Measurement Unitsppb, ppm, μg/m3, mg/m3 (selectable)
 ​Response Time
​< 70 seconds to 95%
 ​Sample Flow Rate
​​800 cc/min ±10%
 Zero Noise​
< 0.02 ppm (RMS)
 Span Noise< 0.5% of reading (RMS) above 5 ppm
 Lower Detectable Limit< 0.04 ppm
 ​Precision
​0.5% of reading above 5 ppm
 ​Linearity
​1% of full scale
 Zero Drift
< 0.1 ppm/24 hours
 Span Drift< 0.5% of full scale/24 hours
 Included I/O1 x Ethernet (TCP/IP)
1 x RS232
2 x front panel USB device ports​
 Optional I/OUniversal Analog Output Board includes (all user-definable):
4 x Isolated Voltage Outputs (5V, 10V; user-selectable)
3 x Individually Isolated Current Outputs (4-20mA)
​Digital I/O Expansion Board includes:
3 x Isolated Digital Input Controls
5 x Isolated Digital Output Controls (user-definable)
3 x Form C Relay Alarm Outputs (user-definable)​
 ​​Weight
​40 lbs (18.1 kg)
 Dimensions (HxWxD)
7” x 17” x 24.3” (178 x 432 x 617 mm)
 ​​Operating Temperature
​0 - 40°C (with US EPA Approval)
 ​​Power
​100V-240V, 50/60 Hz, Typical consumption 115W​​
 Certifications

Specifications are subject to change without notice.
All specifications are based on constant conditions.​
​US EPA: RFCA-1093-093




Downloads

​​​