Model N​200
Chemiluminescence NOX-NO2-NO Analyzer
with NumaView™ Softwar​​​​​e
​​

Description

​The Model N200 uses the proven chemiluminescence detection principle, combined with a state-of-the-art modular architecture, and intuitive operating software to provide accurate and dependable measurements of low-level Nitric Oxide (NO), Nitrogen Dioxide (NO2) and total Nitrogen Oxides (NOx) gases.

The instrument reaction cell uses a diffusion-based inlet nozzle to minimize build-up of contamination on the optics, reducing the frequency of maintenance and minimizing drift. Stability is further enhanced using an automatic baseline ​reference cycle which compensates for any potential baseline drift. The result is sensitive, accurate, and dependable performance under the harshest operating conditions.

Instrument functions and controls are managed through a series of integrated microprocessor-controlled modules utilizing a simple and reliable CAN Bus communications architecture. Each module is independently assembled and calibrated allowing easy and fast field replacement to maximize instrument uptime. The long-life sample filter option further improves efficiency with a ~6 months exchange interval in ambient air quality monitoring applications.

Intuitive operation and calibration of all N Series products is achieved through the NumaView™ Software interface. The graphical user interface (GUI) is customizable, giving the user fast and efficient access to instrument status, as well as measurement data and diagnostic parameters in either numeric or graphical form. NumaView™ Remote Software (included at no charge) provides the same virtual interface and complete instrument control, as well as access to the instrument's large internal data storage buffer from a remote PC or tablet.


Features

  • ​Single or dual range capability
  • Automatic converter efficiency compensation
  • Customizable alerts and continuous self-checking
  • Wide operating temperature range
  • Internal DC-powered pump (optional)
  • Internal zero and span valves (optional)
  • Optional 47mm membrane or long-life​ sample particulate filter​


Specifications


 Measurement Units
​ppb, ppm, μg/m3, mg/m3 (selectable) 
​​​ Response Time
​< 80 seconds to 95% (in switching mode)
 Ranges

Min: 0 - 50 ppb full scale
Max: 0 - 20,000 ppb full scale (selectable, dual-range supported)
 Sample Flow Rate​500 cc/min ±10%
 Zero Noise< 0.1 ppb (RMS)
 Span Noise< 0.2% of reading (RMS) above 50 ppb
 Lower Detectable Limit
< 0.2 ppb

 Precision
​0.5% of reading above 50 ppb

​​ Linearity
​1% of full scale
 Zero Drift
< 0.5 ppb/24 hours
 Span Drift< 0.5% of full scale/24 hours
 Included I/O1 x Ethernet (TCP/IP)
1 x RS232
2 x front panel USB device ports​
 Optional I/O
​Universal Analog Output Board includes (all user-definable):
4 x Isolated Voltage Outputs (5V, 10V; user-selectable)
3 x Individually Isolated Current Outputs (4-20mA)
​Digital I/O Expansion Board includes:
3 x Isolated Digital Input Controls
5 x Isolated Digital Output Controls (user-definable)
3 x Form C Relay Alarm Outputs (user-definable)​
 Weight
Analyzer: 35 lbs (15.9 kg)
External pump: 22 lbs (10 kg)

 ​Dimensions (HxWxD)
​7” x 17” x 24.3” (178 x 432 x 617 mm)
 Operating Temperature0 - 45°C (with US EPA Approval)​

​​​ Power
​Analyzer: 100V-240V, 50/60 Hz, Typical consumption 55W, with internal pump 80W
External pump: 115V, 60 Hz, Typical consumption 85W​
​                             220-240V, 50/60 Hz, Typical consumption 85W

​​​​ Certifications
Specifications are subject to change without notice.
All specifications are based on constant conditions.​​​
​US EPA: RFNA-1194-099​


​​​​

​​​